Advertisement

Early and Mid-Term Durability of Surgeon-Modified and Custom-Made Fenestrated Devices for the Treatment of Complex Aortic Pathology

Published:December 22, 2021DOI:https://doi.org/10.1016/j.avsg.2021.12.006

      Background

      Fenestrated endovascular aneurysm repair (F-EVAR) has allowed successful treatment of patients with complex aortic aneurysms. Custom-made devices (CMDs) are manufactured by companies and tailored to the patient's anatomy to incorporate target vessels, while there is also the “off label” alternative with the devices modified by a surgeon in the operating room: surgeon-modified FEVAR (sm-FEVAR). This study aims to present and compare technical durability of CMDs- and sm-FEVAR for complex abdominal and thoracoabdominal aortic pathologies.

      Methods

      A retrospective cohort study was undertaken including all consecutive patients treated with sm- or CMD-FEVAR during a 3-year period in a single center. Only cases with at least 3 reno-visceral target vessels were included. Primary outcomes were technical success, and freedom from endoleak (EL) (Ia or III; all branch related) and re-intervention during follow-up period. Mortality and morbidity were also recorded.

      Results

      Thirty-two sm-FEVAR patients (81,3% male) and 79 CMD-FEVAR patients (77,2% male) were included. Indication for sm-FEVAR was exclusively urgent, while all CMD- FEVAR were elective. Technical success was similar in sm-FEVAR (100%) and CMD-FEVAR (98,7%) (P = 0.523). Mean follow-up was 16.3 ± 13 and 20 ± 17.3 months for sm-FEVAR and CMD- FEVAR, respectively (P = 0.28). The freedom from EL Ia rate was 91.7% [standard error (SE) 5.7%] at 12 months in sm-FEVAR, while it was 97.7% (SE 2.2%) and 92.3% (SE 4.3%) at 12 and 24 months, respectively in CMD-FEVAR (P = 0.69). The freedom from EL III rate was 95.5% (SE 4.4%) and 88% (SE 8.2%) at 12 and 24 months, respectively in sm-FEVAR, while it was 92.1% (SE 3.8%) and 89.2% (SE 4.7%) at 12 and 24 months, respectively in CMD-FEVAR (P = 0.68). The freedom from re-intervention rate was 91.6% (SE 5.7%) and 84.6% (SE 8.6%) at 12 and 24 months, respectively in sm-FEVAR, while it was 91.7% (SE 4%) and 83.4% (SE 5.9%) at 12 and 24 months, respectively in CMD-FEVAR (P = 0.90). The survival rate was 87.5% (5.8%) and 76.3% (7.9%) at 1 and 12 months, respectively in sm-FEVAR, while it was 93.7% (2.7%) at 1 month in CMD-FEVAR without any other death during FU period. No bridging stent occlusions were noted during follow-up period in any patient.

      Conclusion

      Sm-FEVAR offers good technical success and mid-term clinical outcomes in urgent cases of complex aortic pathologies. Its durability is acceptable and comparable to CMD-FEVAR with a relatively low re-intervention rate.

      Keywords

      To read this article in full you will need to make a payment

      Purchase one-time access:

      Academic & Personal: 24 hour online accessCorporate R&D Professionals: 24 hour online access
      One-time access price info
      • For academic or personal research use, select 'Academic and Personal'
      • For corporate R&D use, select 'Corporate R&D Professionals'

      Subscribe:

      Subscribe to Annals of Vascular Surgery
      Already a print subscriber? Claim online access
      Already an online subscriber? Sign in
      Institutional Access: Sign in to ScienceDirect

      REFERENCES

        • Tsilimparis N
        • Perez S
        • Dayama A
        • et al.
        2nd. Endovascular repair with fenestrated-branched stent grafts improves 30-day outcomes for complex aortic aneurysms compared with open repair.
        Ann Vasc Surg. 2013; 27: 267-273
        • Eagleton MJ
        • Follansbee M
        • Wolski K
        • et al.
        Fenestrated and branched endovascular aneurysm repair outcomes for type II and III thoracoabdominal aortic aneurysms.
        J Vasc Surg. 2016; 63: 930-942
        • Walker J
        • Kaushik S
        • Hoffman M
        • et al.
        Long-term durability of multibranched endovascular repair of thoracoabdominal and pararenal aortic aneurysms.
        J Vasc Surg. 2019; 69: 341-347
        • Oderich GS
        • Ribeiro M
        • Hofer J
        • et al.
        Prospective, nonrandomized study to evaluate endovascular repair of pararenal and thoracoabdominal aortic aneurysms using fenestrated-branched endografts based on supraceliac sealing zones.
        J Vasc Surg. 2017; 65 (1249-59 e10)
        • Sweet MP
        • Hiramoto JS
        • Park KH
        • et al.
        A standardized multi-branched thoracoabdominal stent-graft for endovascular aneurysm repair.
        J Endovasc Ther. 2009; 16: 359-364
        • Fernandez CC
        • Sobel JD
        • Gasper WJ
        • et al.
        Standard off-the-shelf versus custom-made multibranched thoracoabdominal aortic stent grafts.
        J Vasc Surg. 2016; 63: 1208-1215
        • Tsilimparis N
        • Heidemann F
        • Rohlffs F
        • et al.
        Outcome of Surgeon-modified fenestrated/branched stent-grafts for symptomatic complex aortic pathologies or contained rupture.
        J Endovasc Ther. 2017; 24: 825-832
        • Oderich GS
        • Ricotta JJ, 2nd
        Modified fenestrated stent grafts: device design, modifications, implantation, and current applications.
        Perspect Vasc Surg Endovasc Ther. 2009; 21: 157-167
        • Wanhainen A
        • Verzini F
        • Van Herzeele I
        • et al.
        Editor's Choice - European Society for Vascular Surgery (ESVS) 2019 Clinical Practice Guidelines on the Management of Abdominal Aorto-iliac Artery Aneurysms.
        Eur J Vasc Endovasc Surg. 2019; 57: 8-93
        • Safi HJ
        • 3rd Miller CC
        Spinal cord protection in descending thoracic and thoracoabdominal aortic repair.
        Ann Thorac Surg. 1999; 67 (discussion 1953-8): 1937-1939
        • Spanos K
        • Kolbel T
        • Kubitz JC
        • et al.
        Risk of spinal cord ischemia after fenestrated or branched endovascular repair of complex aortic aneurysms.
        J Vasc Surg. 2019; 69: 357-366
        • Spanos K
        • Antoniou GΑ
        • Giannoukas AD
        • et al.
        Durability of fenestrated endovascular aortic repair for juxta-renal abdominal aortic aneurysm repair.
        J Cardiovasc Surg (Torino). 2018; 59: 213-224
        • Kölbel T
        • Spanos K
        • Jama K
        • et al.
        Early outcomes of the t-Branch off-the-shelf multi-branched stent graft in 542 patients for elective and urgent aortic pathologies - a retrospective observational study.
        J Vasc Surg. 2021; S0741-5214 (00916–00912)
        • Eleshra A
        • Hatm M
        • Spanos K
        • et al.
        Early outcomes of t-branch off-the-shelf multibranched stent-graft in urgent and emergent repair of thoracoabdominal aortic aneurysm.
        J Vasc Surg. 2021; S0741-5214: 01980-01987
        • Spanos K
        • Kölbel T
        • Theodorakopoulou M
        • et al.
        Early outcomes of the t-branch off-the-shelf multibranched stent-graft in urgent thoracoabdominal aortic aneurysm repair.
        J Endovasc Ther. 2018; 25: 31-39
        • Scali ST
        • Neal D
        • Sollanek V
        • et al.
        Outcomes of surgeon-modified fenestrated-branched endograft repair for acute aortic pathology.
        J Vasc Surg. 2015; 62: 1148-59 e2
        • Georgiadis GS
        • van Herwaarden JA
        • Antoniou GA
        • et al.
        Systematic review of off-the-shelf or physician-modified fenestrated and branched endografts.
        J Endovasc Ther. 2016; 23: 98-109
        • Dossabhoy SS
        • Simons JP
        • Flahive JM
        • et al.
        Fenestrated endovascular aortic aneurysm repair using physician-modified endovascular grafts versus company-manufactured devices.
        J Vasc Surg. 2018; 67: 1673-1683
        • Canonge J
        • Jayet J
        • Heim F
        • et al.
        Comprehensive Review of Physician Modified Aortic Stent Grafts: Technical and Clinical Outcomes.
        Eur J Vasc Endovasc Surg. 2021; 61: 560-569
        • Juszczak MT
        • Vezzosi M
        • Khan M
        • et al.
        Endovascular repair of acute juxtarenal and thoracoabdominal aortic aneurysms with surgeon-modified fenestrated endografts.
        J Vasc Surg. 2020; 72: 435-444
        • Coles-Black J
        • Barber T
        • Bolton D
        • et al.
        A systematic review of three-dimensional printed template-assisted physician-modified stent grafts for fenestrated endovascular aneurysm repair.
        J Vasc Surg. 2021; 74: 296-306.e1